27.06.2013 Views

USE OF DEKTAK 6M PROFILOMETER - Login | Nanolab, UCLA

USE OF DEKTAK 6M PROFILOMETER - Login | Nanolab, UCLA

USE OF DEKTAK 6M PROFILOMETER - Login | Nanolab, UCLA

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>USE</strong> <strong>OF</strong> <strong>DEKTAK</strong> <strong>6M</strong> PR<strong>OF</strong>ILOMETER<br />

Prepared by Steve Franz 7/17/03<br />

1


Dektak <strong>6M</strong> Profilometer<br />

Introduction:<br />

This spec is an abbreviated procedure on how to operate the Dektak <strong>6M</strong> profilometer located in<br />

<strong>UCLA</strong>'s NRF. A more complete manual is available in pdf format on the Dektak 6 computer (C<br />

drive->Dektak->Docs->Dektak 6) and online. In most cases results can be achieved in more than<br />

one way (eg pull down menus or keyboard commands or icons etc). See full manual for more<br />

details and explanations. Also note that a more capable and similar looking machine, the Dektak<br />

8, is also available. For quick scans and ease of use where stage programmability is not needed,<br />

the Dektak <strong>6M</strong> may be a better choice.<br />

Operating Procedure:<br />

1 Check the log book to verify the last user had no problems.<br />

2 The Dektak <strong>6M</strong> and its computer are usually left on.<br />

a If this is the case, move the mouse to activate the LCD screen. When the<br />

Windows desktop is visible, maximize the D6 program following standard<br />

Windows procedures. This will return you to either the Sample Positioning<br />

window, data window or scan routine window.<br />

b If this is not the case, turn on the Dektak power located to the right of the machine<br />

on the back of the power rack (a bank of LEDs is located on the rack front). Turn<br />

on the computer power located below the Dektak table. If a network login is<br />

requested, double-click the D6 Manual icon to boot up the program. After a<br />

minute or two, the following screen should be seen. Click on the Sample<br />

Positioning Icon (arrow) to bring up the Sample Viewing window.<br />

Fig 1. Main Measurement Screen<br />

2


3 Open the chamber door and verify that the tower and stylus are in the full up position. If<br />

not, raise the stylus and tower by clicking on the<br />

4 Manually move the stage towards you by turning the right stage knob counterclockwise<br />

(see Fig 2).<br />

Fig 2. Wafer stage and controls for Dektak <strong>6M</strong>.<br />

5 Position your sample in the center of the stage. Step to be measured should run<br />

horizontally ie left to right in Fig 2.<br />

Note that sample scanning is done by moving the stage from the rear to the front .<br />

However, this image is rotated on the LCD screen and the stage appears to move from<br />

right to left. You must take this into account when you load your sample.<br />

6 Manually move the stage using the x, y and theta knobs to position your sample under the<br />

probe tip.<br />

7 Making sure that your sample is under the stylus, lower the tower by clicking on the<br />

tower down and null icon .<br />

8 Adjust the lighting by either increasing or decreasing the intensity.<br />

9 Make sure the stylus is NOT touching your sample. Adjust focus by turning the lower<br />

focus ring and adjust magnification by adjusting the upper (numbered) focus ring (Fig<br />

3). Magnification may be adjusted from 70 to 280X. Note: Stylus may be raised and<br />

lowered by pressing the yellow F3 key on the keyboard.<br />

3


Figure 3. Magnification and focus adjust<br />

10 The location where the probe tip will touch your sample is given by the center of the<br />

reticle mark (see below). You will want to position your sample so that the location to be<br />

probed coincides with the reticle center.<br />

Note: If there is misalignment of the reticle mark, see the full procedure in the Dektak<br />

<strong>6M</strong> manual to correct.<br />

Reticle alignment mark<br />

11 Most likely you will want to make some fine adjustments on stage position and rotation.<br />

Proceed as in step 11 WITH THE STYLUS STILL UP.<br />

12 Once the sample is properly positioned under the stylus with the image in focus at the<br />

desired magnification, close the lid on the enclosure and go to the Scan Data Routine to<br />

setup the scan parameters. Click on the scan routine icon to enter this mode.<br />

Note 1: Refer to Dektak <strong>6M</strong> manual for setting up Automated Routines which is a<br />

sequence of single scans.<br />

4


13 The scan data window should appear on the LCD as shown below.<br />

14 Click on the SCAN PARAMETERS button and fill in the white blanks below:<br />

SCAN ID: You can leave this blank.<br />

Scan Length: This is the length of the desired scan in microns and must be from 50-<br />

30,000.<br />

Duration: This is the length of time it takes to complete an entire scan. This number<br />

must be between 3-100 seconds for a maximum of 30,000 data points per scan. Select a<br />

longer duration for large scan lengths or very fine surface roughness where high<br />

horizontal resolution is required. For most users a 10-30 second duration is a good start.<br />

Resolution: This is the horizontal resolution expressed in microns per sample so that a<br />

smaller number gives a higher resolution. Since the total number of points sampled is<br />

fixed, the resolution is related to the scan length and scan duration.<br />

5


Stylus Force: This is the force or effective weight of the stylus in mg. The range is 1.0 to<br />

15 mg. For very soft films or high aspect ratio structures using a nanotip, this number<br />

should be lowered. For the average sample, 15 mg is a good selection.<br />

Measurement Range: The available vertical resolution depends upon the measurement<br />

range selected. When measuring extremely fine geometries, use the most sensitive range<br />

(65 Kang) to get a vertical resolution of 1 Ang. For general applications, the 10 Ang<br />

vertical resolution of the 655 KAng range is usually adequate. When measuring thick<br />

films or very rough or curved surfaces, select the 2620 KAng range with 40 Ang<br />

resolution. Finally for very deep or rough structures the 1 mm range may be chosen.<br />

Profile: Valleys should be used where the area to be measured is below the surface since<br />

90% of the measurement range is provided below the zero. Hills provides 90% above the<br />

zero (as for measuring mesas for example) and hills and valleys provides 50% above and<br />

below the zero line. This is the most commonly used mode (except for the calibration<br />

sample). Click the appropriate profile button to select the desired mode.<br />

15 Click OK to exit the Scan Parameters Dialog Box.<br />

16 Click the Display Parameters button in the Scan Routine Window. A Dialog box should<br />

appear as below:<br />

17 If you want the software leveling to be automatic, check the automatic leveling box and<br />

choose a cursor width (20-50 microns is a good start). You must also know the cursor<br />

position, which can be obtained from the data window. Enter these numbers in the cursor<br />

positions box for both the R (reference) cursor and M (measurement) cursor. If you are<br />

unsure of how to do this, uncheck the automatic leveling and level manually as described<br />

below.<br />

18 Choose automatic ranging unless you want to manually set the vertical range of the<br />

graph. Click OK to exit the Display Parameters Box.<br />

19 If Data Processing filtering is desired, see Dektak <strong>6M</strong> manual.<br />

20 Click on the “scan” icon to initiate the scan.<br />

21 Assuming the Graphics & Video is selected , the stylus should automatically lower<br />

to the set location and the stage should begin moving. A video image of the stylus as well<br />

as a data screen should be displayed. When the scan is completed, the stylus lifts up, and<br />

6


the stage returns to where the scan originated. A data plot screen similar to below should<br />

be seen on the LCD monitor.<br />

22 In the example above, the plot is obviously not level and so manual leveling needs to be<br />

done. To do so:<br />

A Set the default cursor bandwidth (Plot-->Default Bands) from the pull-down<br />

menu.<br />

B Move the R cursor by using the mouse and clicking to grab and drag the cursor to<br />

one of the reference positions to be leveled. Alternately you can select the R<br />

button (bottom right) and click the appropriate arrow button.<br />

C Similarly, move the M cursor to the 2 nd leveling location. Try to keep the cursors<br />

as far apart as possible for best results.<br />

D Level the plot from the pull-down menu ( Plot _--->Level).<br />

Note 1: If automatic leveling was selected in the Display Parameters Box (step 17) AND<br />

BOTH cursors were properly set, manual leveling should not be necessary.<br />

Note 2: If the out-of-level is very severe such that manual leveling still does not produce<br />

an acceptable trace, mechanical leveling may be necessary and is covered later in the D6<br />

manual.<br />

7


23 The zero point of the trace for this plot may now be set, if desired by moving the R cursor<br />

as described above and then using the pull-down menu (Plot -->Zero ). This will shift the<br />

curve up or down to make the R cursor intercept 0.<br />

24 A quick step height reading can be made by positioning the M cursor at the top of the<br />

step plot and the R at the bottom and then reading the number under Vert D at the bottom<br />

right. HOWEVER it is recommended that you use the delta average step height<br />

measurement for more accurate results. To do this:<br />

A Position the M cursor at the top of the step scan and the R cursor at the bottom.<br />

B Using the pull-down menu select Analysis-->Analytical Functions.<br />

C Under Height select ASH for average step height function<br />

D Click the Measure field at the bottom of the dialog box. If you wish to add the<br />

function to all scan routines select Edit--> Append Function while in the Scan<br />

Routine Window.<br />

E Click the COMPUTE button to calculate the average step height. The result is<br />

displayed in the area to the left of the data pane.<br />

Note: The average is calculated using the default cursor bandwidths selected earlier in<br />

step 22. If the average is to be taken over a wider or narrower plot range, select<br />

Plot-->Bandwidths and enter the number in microns which is desired and then recompute<br />

the average step height as previously described.<br />

25 If desired, a portion of the scan can be re-plotted at a magnified value. Using the mouse,<br />

click and drag over the area to be magnified in the data plot window. When the selection<br />

box turns green, it has enough data to replot. Click the mouse a 2 nd time to set the range.<br />

Select Plot --> Replot from the pull-down menu. Select Plot --> Replot to return to the<br />

original trace. Up to 9 plot boundaries may be saved for a given trace as described later in<br />

the D6 manual.<br />

26 The NRF’s Dektak <strong>6M</strong> is equipped with both a floppy and a CD ROM Burner. After the<br />

plot is formatted to your taste you can save it to disk by selecting File-->Save as and<br />

navigating to the appropriate drive following standard windows procedure.<br />

27 If more scans on the same sample are desired, go back to the sample positioning window<br />

by selecting Window-->Sample Positioning from the pull-down menu. With the stylus<br />

up, manually reposition the stage and refocus and readjust magnification if necessary.<br />

Continue to follow the above procedures to obtain a new scan.<br />

28 When you are through with your sample, press the tower up icon, and manually<br />

move the stage out to unload your sample. Remove your sample and position your next<br />

sample on the center of the stage if desired and move the stage back under the stylus.<br />

Continue as in steps 7 to 26 above.<br />

8


29 If you are through using the Dektak, make sure the stage is centered and the tower is up.<br />

Turn off the illumination using the dim icon , close the chamber door and minimize<br />

the D6 Manual program. It is not necessary to shut down Windows or the computer.<br />

30 Fill out the log book completely. Don’t forget to remove your floppy or CD if you<br />

brought one.<br />

9

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!