ESPaDOnS: spectral response and global efficiency http://webast.ast.obs-mip.fr/magnetisme/espadons_new/efficiency.html<br />

radiation corresponds to a temperature of 1500K (dashed line)<br />

or 2000K (dash-dot line). The curve we measure is in good<br />

agreement with the 1500K prediction as far as flux is<br />

concerned, while it agrees better with the 2000K expectations as<br />

far as spectral response is concerned (presumably because the<br />

halogen lamp used does not behave like a pure blackbody). The<br />

signal to noise ratio we obtain at 550nm (independent on lamp<br />

colour, by definition of V magnitudes) is larger than expected<br />

by about 40% (presumably due to uncertainties in the<br />

calibration of the guiding channel).<br />

In any case, it indicates that the instrument throughput is<br />

nominal within about a factor of 2. Tests on the sky (using stars<br />

of known brightness and color) are needed to improve the<br />

accuracy of this estimate.<br />