CHANGING ROTORS<br />
Lift the retaining clips with a finger or thumb and slide back about 1 cm. The<br />
Teflon lid can now be removed. Hold the shaft collar stationary when screwing<br />
the endplate in or out. Alternatively and more conveniently, turn the shaft collar<br />
and hold the endplate stationary.<br />
1 Unscrew and remove endplate. → 2 Slide rotor up with finger → 3 Remove rotor.<br />
and place shoehorn under<br />
lower edge. ↓<br />
6 Use endplate in center hole 5 Slide rotor up with finger 4 Lower new rotor<br />
to slide rotor up into position. to allow removal of shoehorn. into position.<br />
Screw in endplate by hand only. ← ←<br />
A tight endplate can be levered loose with the endplate key. Insert into one of<br />
the holes at the side of the endplate. See diagram page 3.<br />
If the slack between the screw threads on the endplate and the hole in the rotor<br />
causes slight eccentricity, loosen the endplate and screw up in a new position<br />
relative to the rotor. Repeat until concentric. Eccentricity of the sorbent layer,<br />
or the glass rotor, does not affect performance.<br />
SORBENT LAYER THICKNESS<br />
1 mm Layers. For separation of small samples, up to 100 mg per component,<br />
250 mg total sample.<br />
2 mm Layers. For larger sample loads, up to 300 mg per component, 750<br />
mg total sample. Also for small samples of low solubility or which tend to tail.<br />
The resolving powers of 1 and 2 mm layers are not significantly different for<br />
light loadings.<br />
4 mm Layers. For very large loads, up to about 1.5g. Gain experience with 1<br />
and 2 mm layers before using these thick layers.<br />
Layers up to a thickness of 8 mm can be prepared but are troublesome to use.<br />