Chromatotron
Chromatotron
Chromatotron
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
_________________________________APPENDICES<br />
APPENDIX 1. THE TEST MIXTURE<br />
A mixture of high Rf colored compounds is useful for testing sorbent layers.<br />
"The test mixture" refers to a mixture of the 2,4-dinitrophenylhydrazones of<br />
cyclopentanone and cycloheptanone. The same derivatives of open-chain<br />
ketones are also suitable. Azobenzene derivatives and the commercial dye<br />
mixtures used with regular TLC plates are not recommended since they are<br />
not completely stable and contain low Rf impurities that remain on the sorbent.<br />
APPENDIX 2. INTRODUCING LARGER VOLUMES OF SAMPLES<br />
Large volumes of sample solutions<br />
can be introduced through a syringe<br />
barrel clamped at the same height as<br />
the Teflon funnel.(Suggested by<br />
J.G. Krause, Niagara University).<br />
A cotton plug filter is required<br />
in the syringe barrel to prevent<br />
clogging of the stainless steel<br />
capillary tubing by solid impurities.<br />
41