PROBLEM CAUSE CURE<br />

Solvent leaks from pump or tube<br />

fittings.<br />

1 Increased pressure due to blockage<br />

on output side of pump.<br />

2 Flared ends of tubing damaged.<br />

Disconnect tubing and inlet.<br />

Check for blockages. Do not<br />

tighten with tools! Hand tighten<br />

only. Thread tape is not<br />

effective.<br />

Pump does not start. Unplug pump, remove cover<br />

(thumbscrew at rear of pump).<br />

Take off drive belt, turn pulley<br />

by hand to determine if problem<br />

is in pump head or motor.<br />

Check belt tension.<br />

Solvent inlet becomes tight in<br />

lid.<br />

Particles of sorbent on Teflon<br />

lid.<br />

<strong>Chromatotron</strong> oscillates from<br />

side to side.<br />

1 Inlet has swelled through<br />

absorption of solvents.<br />

2 Deposits on threads.<br />

1 Sorbent dust not blown off after<br />

scraping rotor.<br />

2 Excessive nitrogen flow.<br />

1 Rotor unbalanced due to uneven<br />

evaporation of solvent.<br />

2 Unbalanced home made glass rotor.<br />

Rotors crack at center. Rotor support collar and shaft collar<br />

loose.<br />

Endplate becomes etched.<br />

Flaky patches in main vessel.<br />

<strong>Chromatotron</strong> produces a<br />

rubbing sound that continues<br />

after switching off, until the<br />

instant the rotor stops.<br />

Corrosion by strong acids, usually<br />

HCl impurity in chloroform.<br />

Deposits on felt seal causing adhesion<br />

to the rotor support collar.<br />

Motor does not start. Rotor support collar in contact with<br />

main vessel.<br />

Elinco motor has insufficient<br />

torque. Slow start up. May run<br />

at half speed.<br />

Allow inlet and lid to dry out for<br />

24 hr.<br />

Clean with solvents or scrape<br />

off deposits.<br />

Rethread with 5/16-18 tap and<br />

die.<br />

Blow off sorbent after scraping.<br />

Wipe lid with moist paper towel.<br />

Evaporate solvent from rotor<br />

with nitrogen before switching<br />

off.<br />

Remove rotor and allow solvent<br />

to evaporate.<br />

See Replacing the Rotor<br />

Support Collar, page 34.<br />

Add solid sodium bicarbonate<br />

or alumina to chloroform. Test<br />

with wet pH paper.<br />

Use dichloromethane in place<br />

of chloroform.<br />

No maintenance required<br />

unless adhesion prevents start<br />

up of motor. See the Felt Seal,<br />

page 34.<br />

See Adjusting the Main Vessel,<br />

page 34.<br />

Low supply voltage. Use motor capacitor 1 MFD<br />

larger than specified or<br />

increase supply voltage with<br />

variable transformer.<br />

Shoehorn or endplate key lost. Use a small hex key (Allen<br />

wrench).<br />

39