For recycle, proceed as in normal elution A until the bands to be recycled are<br />

about 1 cm from the edge of the sorbent then remove the solvent input tube<br />

from the solvent reservoir, B. After waiting about l0 sec to reduce the amount<br />

of solvent in the pump and <strong>Chromatotron</strong>, connect the filter (with cotton) to the<br />

output tube (with detachable part removed),C. The pump will recycle solvent<br />

from the <strong>Chromatotron</strong> for about l min then both solvent and vapor will recycle.<br />

A B C<br />

If the l0 sec delay at stage B is omitted, solvent may back up slightly in the<br />

collection channel causing some remixing of bands. The solvent flow rate must<br />

be held constant for some time before recycle commences to prevent surges<br />

of solvent flow.<br />

When the bands of interest have been returned to the rotor it is preferable to<br />

return to mode A since in mode C evaporation will slowly reduce the amount of<br />

solvent available for recycle.<br />

Bands will be discouragingly diffuse when newly recycled back to the sorbent.<br />

However, on further development they will sharpen up considerably. After one,<br />

two or three stages of recycle, depending on the loading and the tailing<br />

tendency of the particular compounds, the bands will become so broad that<br />

further recycle will cause the head of one to catch up with and overlap the tail<br />

of another. At this stage the band or bands should be eluted or if recycle is to<br />

continue they must be sharpened by multiple development (page 18). If the Rf<br />

of the bands is 0.5 or more, multiple development is very effective for band<br />

sharpening, allowing an unlimited number of recycles to be performed.<br />

CONNECTING CHROMATOTRONS IN SERIES<br />

Resolving power and capacity can be increased by connecting two (or more)<br />

<strong>Chromatotron</strong>s in series. Pump the output of one to the input of the next.<br />

Back-up of solvent in the first <strong>Chromatotron</strong> can be prevented by setting the<br />

intermediate pump at a higher flow rate so that both solvent and some vapor<br />

are pumped.<br />

19