If the RS485 interface was configured as Modbus master, the following screen will be displayed. You should set up a tag<br />
(name - maximum of 16 characters), unique for each channel, and optionally a unit (maximum of 10 characters) for the<br />
reading value.<br />
In addition, for each channel, you should set up which network slave is to be read, what Modbus command sholud be<br />
used and what is the initial register to be read from the slave. In the current version, the number of registers is always<br />
"1", which means that you can not do block reading.<br />
An error value must also be configured. This value is shown on the channel when there is a problem, for example, a<br />
communication error with the slave.<br />
At the end of the channel configuration, you must click on the “Add” button in order to enter the channel in the list to the<br />
left. If you want to remove a channel from the list, you must select it from the list and click on "Remove".<br />
Remote Channels - Setting up a channel<br />