VISIFERM DO Operating Instructions<br />
configured the ECS interface completely. The above values<br />
correspond to a typical OXYFERM oxygen sensor with 60 nA at air<br />
and at a temperature of 25°C.<br />
HINT: If you input ‘256’ instead of ‘512’ in the 2nd step, then the current<br />
corresponding to the value in the field ‘Fixed test value’ is supplied over the<br />
ECS interface! This would be 80 nA in our example. This test mode is helpful<br />
in order to test the ECS interface, the cable and the oxygen measuring device!<br />
Activation of the 4-20 mA mode<br />
Configure the sensor with VisiConfigurator as follows:<br />
1. At first, it is essential to adjust the right measuring value (vide infra)!<br />
In the following we assume that you have decided for the<br />
measurement in %-oxygen air saturation (%-sat). The adjustments<br />
for the other measured values take place in a similar way.<br />
2. If you are not yet on the specialist level already: User Level /<br />
Command: Change user level / New user level: S / Password: see<br />
table above / Go. Now a ‘S’ has to appear on the right side in the<br />
field ‘New user level’. Thus you have the user rights of a specialist<br />
and may reconfigure the sensor!<br />
3. Configuration Interface / Command: Write current interface mode /<br />
Interface: 2 / Go. Now a ‘2’ has to appear on the right side of the field<br />
‘Interface’. Thus you have deactivated the ECS interface and<br />
activated the 4-20 mA interface. Now the 4-20 mA interface has to be<br />
adjusted!<br />
4. As an example, 4 mA 0% and 20 mA 160% oxygen air saturation<br />
shall be assigned. Configuration Interface / Command: Write<br />
configuration 4…20 mA interface / Oxygen value at 4mA: 0 / Oxygen<br />
value at 20mA: 160 / Fixed test value: 10 / Go. Now the same values<br />
have to appear again on the right side. Furthermore, the measuring<br />
unit is displayed for control. Thus you have configured the 4-20 mA<br />
interface completely.<br />