VISIFERM DO Operating Instructions<br />

configured the ECS interface completely. The above values<br />

correspond to a typical OXYFERM oxygen sensor with 60 nA at air<br />

and at a temperature of 25°C.<br />

HINT: If you input ‘256’ instead of ‘512’ in the 2nd step, then the current<br />

corresponding to the value in the field ‘Fixed test value’ is supplied over the<br />

ECS interface! This would be 80 nA in our example. This test mode is helpful<br />

in order to test the ECS interface, the cable and the oxygen measuring device!<br />

Activation of the 4-20 mA mode<br />

Configure the sensor with VisiConfigurator as follows:<br />

1. At first, it is essential to adjust the right measuring value (vide infra)!<br />

In the following we assume that you have decided for the<br />

measurement in %-oxygen air saturation (%-sat). The adjustments<br />

for the other measured values take place in a similar way.<br />

2. If you are not yet on the specialist level already: User Level /<br />

Command: Change user level / New user level: S / Password: see<br />

table above / Go. Now a ‘S’ has to appear on the right side in the<br />

field ‘New user level’. Thus you have the user rights of a specialist<br />

and may reconfigure the sensor!<br />

3. Configuration Interface / Command: Write current interface mode /<br />

Interface: 2 / Go. Now a ‘2’ has to appear on the right side of the field<br />

‘Interface’. Thus you have deactivated the ECS interface and<br />

activated the 4-20 mA interface. Now the 4-20 mA interface has to be<br />

adjusted!<br />

4. As an example, 4 mA 0% and 20 mA 160% oxygen air saturation<br />

shall be assigned. Configuration Interface / Command: Write<br />

configuration 4…20 mA interface / Oxygen value at 4mA: 0 / Oxygen<br />

value at 20mA: 160 / Fixed test value: 10 / Go. Now the same values<br />

have to appear again on the right side. Furthermore, the measuring<br />

unit is displayed for control. Thus you have configured the 4-20 mA<br />

interface completely.<br />

page 15 of 48