3. Safety instructions<br />

VISIFERM DO Operating Instructions<br />

This sensor only is to be used for the intended use and in an optimum safety<br />

and running conditions. The specifications such as temperature or pressure<br />

being defined in the chapter ‘Technical data’ may not be exceeded under any<br />

circumstances. Threats are imminent if the sensor is not operated correctly or<br />

appropriately.<br />

Assembly and maintenance only have to be performed by trained personnel.<br />

Be careful that the PG 13.5 thread and O-ring are not damaged during the<br />

screwing in into the process. O-rings are wearing parts which regularly have<br />

to be exchanged, at the latest after one year. Even where all necessary safety<br />

measures have been complied with, there are still further risks involving leaks<br />

or mechanical damage to the armature. Where there are seals or screws,<br />

gases or liquids can leak out undetected.<br />

Before removing the sensor make sure that any process medium cannot be<br />

withdrawn.<br />

4. Initial operation<br />

Control the VISIFERM DO sensor during unpacking according to possible<br />

defects. Rejected <strong>sensors</strong> are to be sent to your HAMILTON merchant in the<br />

original packaging.<br />

1 2<br />

3 4<br />

5 6<br />

1: socked head, 2: key areas with serial number, 3: PG 13.5 thread, 4:<br />

O-ring, 5: sensor shaft with fusion number of the stainless steel, 6, 7:<br />

sensor cap with oxygen sensitive silicone luminophore (sensory element),<br />

serial number of the sensor cap and fusion number of the stainless steel.<br />

VISIFERM DO <strong>sensors</strong> are delivered by the factory according to the<br />

operation with the integrated 4-20mA power interface as well as to the always<br />

active serial RS-485 bus interface (Modbus RTU – protocol). You also find this<br />

information on the provided certificate which also contains the serial number<br />

as well as the most important specifications.<br />

ATTENTION !<br />

For the operation at measuring amplifiers which are designed for<br />

classical amperometric <strong>sensors</strong> (Clark cells) a software<br />

configuration in the ECS mode (electrochemical sensor) is needed!<br />

page 5 of 48<br />