VISIFERM DO Operating Instructions<br />
10. Testing and maintenance<br />
Calibration<br />
According to the standards, the sensor can be calibrated at air or in oxygenfree<br />
environment (zero point). This is very rarely necessary in the most<br />
applications. If there are already displacements of measuring values of only<br />
few percents within few days, then a damage of the luminophore due to<br />
measuring medium and cleaning medium has to be expected. In this case it<br />
has to be considered whether a more frequent exchange of the sensor cap<br />
can be accepted or whether another measuring principle e.g. with a classical<br />
OXYFERM sensor should be used!<br />
HINT: The sensor needs a preheating time of 10-15 minutes after switching<br />
on. In this phase measurements are already possible. However, for an optimal<br />
calibration you should wait this time.<br />
Calibration in Modbus or 4-20 mA mode<br />
The calibration of the sensor in the 4-20 mA mode only is possible via Modbus<br />
RTU. Therefore, most easily VISICAL or the VisiConfigurator freeware with<br />
the USB-RS-485 Modbus converter can be used (see chapter ‘Configuration<br />
of the <strong>sensors</strong>‘ and ‘Accessories‘). The operation of the accessories is shown<br />
by the specific instructions.<br />
During the calibration the sensor independently examines the correctness and<br />
stability of the oxygen signal.<br />
Calibration in ECS mode<br />
The calibration in ECS mode takes place like in classical <strong>sensors</strong> in the<br />
measuring device at which the sensor is connected. For this, consult the<br />
operation instructions of the measuring device!<br />
HINT: The concept of VISIFERM DO <strong>sensors</strong> enables the calibration of all<br />
used <strong>sensors</strong> in the lab with a VISICAL module or with a personal computer or<br />
laptop via Modbus and then a direct use at classical measuring devices<br />
without a new calibration at the measuring device. This is possible, since with<br />
page 19 of 48