VISIFERM DO Operating Instructions<br />

10. Testing and maintenance<br />

Calibration<br />

According to the standards, the sensor can be calibrated at air or in oxygenfree<br />

environment (zero point). This is very rarely necessary in the most<br />

applications. If there are already displacements of measuring values of only<br />

few percents within few days, then a damage of the luminophore due to<br />

measuring medium and cleaning medium has to be expected. In this case it<br />

has to be considered whether a more frequent exchange of the sensor cap<br />

can be accepted or whether another measuring principle e.g. with a classical<br />

OXYFERM sensor should be used!<br />

HINT: The sensor needs a preheating time of 10-15 minutes after switching<br />

on. In this phase measurements are already possible. However, for an optimal<br />

calibration you should wait this time.<br />

Calibration in Modbus or 4-20 mA mode<br />

The calibration of the sensor in the 4-20 mA mode only is possible via Modbus<br />

RTU. Therefore, most easily VISICAL or the VisiConfigurator freeware with<br />

the USB-RS-485 Modbus converter can be used (see chapter ‘Configuration<br />

of the <strong>sensors</strong>‘ and ‘Accessories‘). The operation of the accessories is shown<br />

by the specific instructions.<br />

During the calibration the sensor independently examines the correctness and<br />

stability of the oxygen signal.<br />

Calibration in ECS mode<br />

The calibration in ECS mode takes place like in classical <strong>sensors</strong> in the<br />

measuring device at which the sensor is connected. For this, consult the<br />

operation instructions of the measuring device!<br />

HINT: The concept of VISIFERM DO <strong>sensors</strong> enables the calibration of all<br />

used <strong>sensors</strong> in the lab with a VISICAL module or with a personal computer or<br />

laptop via Modbus and then a direct use at classical measuring devices<br />

without a new calibration at the measuring device. This is possible, since with<br />