FIGURE 1.5 INPUT VOLTAGE/CURRENT/POWER RELATION DIAGRAM<br />
The theory of the CP operation comes from the divider circuit built in the load. Because of the nonlinear<br />
characteristic of the divider, the input voltage, input current, and power setting will limit the CP mode of operation.<br />
The CP mode will not allow the load to operate beyond the full range of the Constant Power setting.<br />
AIRFLOW FUNCTION<br />
Airflow plays and important part in power dissipation. The programming electronic load uses cooling fans as a<br />
standard cooling system for removing the heat generated from the power module.<br />
Do not impede or obstruct the ventilation holes for any reason to prevent the power components from overheating.<br />
All <strong>PEL</strong>’s have OTP protection to prevent the unit from overheating. The standard programming electronic loads<br />
have forced air-cooling fans installed as the main cooling system and operate when the power-handling unit is turned<br />
on.<br />