TOGGLE MODE:<br />
In toggle mode, the <strong>PEL</strong> input alternates between two user-defined input levels. Unlike transient-continuous<br />
mode, the input only changes from one level to another, one transition at a time.<br />
Transient<br />
Level<br />
Main<br />
Level<br />
Toggle<br />
t<br />
Typical Application: Testing the transient recovery time for a CV power supply. The recovery time for a<br />
power supply is defined as the time required for the supply to settle to a predefined settling band following a<br />
change in load current caused by an induced transient (pulse).<br />
LCL Operation:<br />
• From keypad select operating mode (CC, CV, CR, CP)<br />
• Select transient operation for toggle mode.<br />
• Set MAIN and TRANSIENT levels by pressing the “MAIN” and “TRANS” keys.<br />
• To begin the toggle operation, turn on the input and press the “TR ON/OFF. The LCD will<br />
display a flashing “TG” to indicate that the <strong>PEL</strong> is in toggle mode.<br />
All the operating modes can be programmed via the front panel or by way of the GPIB/RS232 Interface.<br />
In addition, each transient can be triggered through an external TTL compatible signal.<br />