6.5.2 Tab „Inclination“<br />
• Select <strong>IMP</strong> > Real-time view (available only if an <strong>IMP</strong> Unit is connected).<br />
• Select the tab Inclination.<br />
The following screen is displayed.<br />
On the left side, the force actually acting on the sensors is shown. Is the <strong>IMP</strong> Unit<br />
positioned on its underside (e.g. like image right), so gravity acts on the z sensor with<br />
-1.0g.<br />
Since no force is acting on x and y sensors in resting position, the chart shows exact<br />
vertical, perpendicular position (red dot).<br />
Further information ➜ Test Shake Sensors, page 24.<br />
INHECO 38