hydrosulfite dissolved in aqueous media will<br />

consume O 2 rapidly, and can be used to prepare a<br />

0% concentration. Air-saturated values for<br />

various solvents and salt solutions can be found<br />

in textbooks.<br />

Calibration Data<br />

Once you have calibrated your sensor system, the<br />

calibration data is stored in two files. It is stored in<br />

the OOISensors.cfg file, which is the application<br />

configuration file. The calibration data is called from<br />

this binary file each time you use your sensor system<br />

and software.<br />

Calibration data is also stored in an ASCII file (or text<br />

file) so that you use read the data and even import it<br />

into other application programs such as Microsoft Word<br />

and Excel. This ASCII file is called chXFoxy.cal,<br />

where "X" stands for the spectrometer channel ("0" for<br />

master spectrometer, "1" for spectrometer channel 1,<br />

"2" for spectrometer channel 2 and so on). The<br />

chXFoxy.cal file is not used by the OOISensors<br />

application; it is strictly for analyzing calibration data.<br />

(If you have temperature data in this file, temperature<br />

will be displayed as Kelvin.)<br />

Re-calibration<br />

FOXY probes do need re-calibration from time to<br />

time. The following factors necessitate re-calibration:<br />

1. If you use a FOXY probe in a harsh environment<br />

that degrades the probe coating.<br />

2. If you expose a FOXY probe to the LED source<br />

for long periods of time, causing faster<br />

photobleaching of the ruthenium compound.<br />

3. If you re-condition your FOXY probe.<br />

4. If you sterilize your FOXY probe with processes<br />

such as autoclave or gamma radiation.<br />

30