Positive and Negative Pressure Areas<br />
All positive pressure plenums in the cabinet, the work area and the unipressure flow system, are<br />
surrounded by negative pressure. If a leak should occur in the positive pressure plenum, the<br />
suction from the blower will capture the air from the leak and recirculate it. Also, if a leak<br />
should occur in a contaminated negative pressure plenum, such as the sidewall or rear wall, the<br />
suction will pull air in through the leak. Thus contaminated air in a negative pressure plenum<br />
will not leak out into the room.<br />
8