Controls<br />
Controls and alarms are located on the front panel of the unit.<br />
Duplex Outlet<br />
Switch<br />
Fluorescent Light<br />
Switch<br />
UV Light<br />
Switch<br />
Motor/Blower<br />
Switch<br />
Indicator lights<br />
below each switch<br />
Alarm Reset Switch<br />
SEE NOTE BELOW<br />
Fig. 2: Control Switches<br />
Alarm Reset Switch<br />
For normal operation, the bottom edge of the view screen, also called the “sash”, is set to<br />
restrict the front opening to 200 mm (8”). The sash alarm will sound whenever the sash is<br />
above this position. Pressing the alarm-reset button will mute the alarm for five minutes.The<br />
light will continue to flash.<br />
NOTE: Per EN Standard 12469:2000 this fuction is not operational. The sash alarm will sound<br />
continually until it is returned to its proper level.<br />
Motor/Blower Switch<br />
This switch turns the blower on and off. The fluorescent light will not work unless the blower is<br />
running.<br />
Duplex Outlet Switch<br />
This switch turns the outlets on each side of the work area on and off.<br />
Fluorescent Light Switch<br />
This switch turns the fluorescent light on and off. The fluorescent light will not work unless the<br />
blower is running.<br />
UV Light Switch<br />
Turns the UV light on and off. The UV light will not work unless the view screen is fully<br />
closed and the fluorescent light is off.<br />
5