Fill the specimen Dewar with liquid nitrogen before freezing.<br />

Open the specimen Dewar drawer by pulling the drawer away from the machine.<br />

Position the LN 2 filled specimen Dewar firmly against the metal holder and then close<br />

the drawer.<br />

Specimen LN Dewar<br />

Metal holder<br />

When the system is fully cooled (approximately 30 minutes after the internal LN 2<br />

pump starts) and the small LN 2 Dewar is filled, the main screen shows “ready for<br />

freezing” in the top left corner. The freezing process can begin.<br />

5.2.3 Inserting the specimen holder in the loading station<br />

The loading mechanism must be open to insert the three-piece cartridge consisting of<br />

an upper cylinder, carrier plate and lower cylinder.<br />

Open the loading mechanism by rotating it towards you into the front locking position<br />

with the loading lever.<br />

Hold the bottom half of the closure with the right forefinger and turn the top half to a<br />

vertical position with the loading lever.<br />

Front locking<br />

position of the<br />

loading<br />

mechanism<br />

Loading lever<br />

Page 35 <strong>Leica</strong> <strong>EM</strong> <strong>HPM100</strong> Operating Manual Version 12/10