Operation Manual<br />
<strong>Helena</strong> C-4 - Software C11.11b<br />
<strong>Helena</strong> Biosciences Europe<br />
UK<br />
5.1.8. Patient Identification<br />
Four choices are available:<br />
• No Patient ID<br />
• Extra Input<br />
• Autoseries<br />
• Barcode<br />
Use the cursor keys to select desired mode, press “Enter” or “Menu” to advance. If No<br />
Patient ID is selected, results will be printed out without a patient identification number. If<br />
Extra Input is chosen, the user enters a patient identification number when running each<br />
test (in the Analysis mode). The third option, Autoseries, allows the user to enter a starting<br />
patient identification number. Each sample run (in the Analysis mode) is then automatically<br />
incremented by one from the starting patient identification number entered by the user. If<br />
Barcode is active, an alphanumeric Pat.Id. is entered by a barcode-scanner. No manual<br />
input or correction is possible. The maximum length of the Pat.Id. in that mode is 20<br />
characters. Limited by the space on the LCD, only the first 10 characters will be displayed.<br />
5.1.9. Date/Time<br />
U.S. date format (mm/dd/yy) will be selected if “English” is chosen as the language. The<br />
date format is dd/mm/yy if German is selected. The Time format is not changeable.<br />
To return to the main menu, press “Menu”.<br />