Operation Manual<br />

<strong>Helena</strong> C-4 - Software C11.11b<br />

<strong>Helena</strong> Biosciences Europe<br />

UK<br />

5.2. Setup Test<br />

The specific parameters for each test are entered in this menu. Once the initial data is<br />

entered and saved it is not necessary to go into this menu for routine testing. The submenu<br />

of “SETUP TEST” is illustrated in Figure 10. To enter this submenu, press #2 from the<br />

main menu. The default values for “SETUP TEST” are:<br />

METHOD:<br />

UNITS:<br />

CALIBRATION CURVE:<br />

AUTOSTART:<br />

CLOT (FIB=Clauss; DD+vWF=IMMUN;<br />

PLG+a2AP=Kinetic)<br />

seconds or dE/Min<br />

Reset to zero<br />

500 (FIB=250; DD=1000)<br />

5.2.1. Setup Test<br />

To select a test enter the numeric code of the designated test. Alternatively, the Up/Down<br />

arrow keys can be used to scroll through the entire test menu. For example, key in #1 to<br />

select PT. Press “Enter” to confirm selection. The “METHOD” can also changed when the<br />

area on the screen is ‘blinking’.<br />

TEST:<br />

PT<br />

METHOD: CLOT<br />

AUTOSTART: 500<br />

Depending which test is active, following methods can be selected:<br />

• CLOT Clotting assays<br />

• CLOT+FIB Fibrinogen will be derived from PT<br />

• CLAUSS Fibrinogen according to CLAUSS<br />

• KINETIC Chromogenic Assay<br />

• 100mOD Chromogenic Ecarin Assays<br />

• IMMUNO Immunoturbidimetric Assays<br />

23