Operation Manual<br />
<strong>Helena</strong> C-4 - Software C11.11b<br />
<strong>Helena</strong> Biosciences Europe<br />
UK<br />
5.2. Setup Test<br />
The specific parameters for each test are entered in this menu. Once the initial data is<br />
entered and saved it is not necessary to go into this menu for routine testing. The submenu<br />
of “SETUP TEST” is illustrated in Figure 10. To enter this submenu, press #2 from the<br />
main menu. The default values for “SETUP TEST” are:<br />
METHOD:<br />
UNITS:<br />
CALIBRATION CURVE:<br />
AUTOSTART:<br />
CLOT (FIB=Clauss; DD+vWF=IMMUN;<br />
PLG+a2AP=Kinetic)<br />
seconds or dE/Min<br />
Reset to zero<br />
500 (FIB=250; DD=1000)<br />
5.2.1. Setup Test<br />
To select a test enter the numeric code of the designated test. Alternatively, the Up/Down<br />
arrow keys can be used to scroll through the entire test menu. For example, key in #1 to<br />
select PT. Press “Enter” to confirm selection. The “METHOD” can also changed when the<br />
area on the screen is ‘blinking’.<br />
TEST:<br />
PT<br />
METHOD: CLOT<br />
AUTOSTART: 500<br />
Depending which test is active, following methods can be selected:<br />
• CLOT Clotting assays<br />
• CLOT+FIB Fibrinogen will be derived from PT<br />
• CLAUSS Fibrinogen according to CLAUSS<br />
• KINETIC Chromogenic Assay<br />
• 100mOD Chromogenic Ecarin Assays<br />
• IMMUNO Immunoturbidimetric Assays<br />
23