Operation Manual<br />

<strong>Helena</strong> C-4 - Software C11.11b<br />

<strong>Helena</strong> Biosciences Europe<br />

UK<br />

5.3.3. Duplicate testing<br />

If duplicate testing is desired, enter the same patient identification number for channels 1 &<br />

2 or 3 & 4 twice. The mean result will automatically be printed along with the individual<br />

channel data.<br />

<br />

<br />

If the duplicates differ more than 7.5% from the mean<br />

value, the mean result will be flagged with ‘X’ !<br />

Pressing key “0” during optic activation, will activate all<br />

channels in duplicate mode<br />

5.3.4. Starting the Analysis<br />

Place the required number of cuvettes in the cuvette pre-warming positions. To continue<br />

with the PT example, pipette 25 µL of sample to each cuvette well. Press the “Timer 1”<br />

button to start the stopwatch. Place the Thromboplastin / CA mixed reagent with magnetic<br />

stir bar in the center large reagent pre-warming position.<br />

Transfer the first cuvette with sample to the measurement position. Once patient<br />

identification numbers have been entered, press the Optic keys again to activate the<br />

channels.<br />

FIB: ACTIVE 01:30<br />

FIB: ACTIVE 00:26<br />

FIB: 00:00<br />

FIB: 00:00<br />

When the timer has reached 1 minute, pipette 50 µl of mixed Calcium Chloride and<br />

Thromboplastin reagent into the first cuvette position. If autostart is enabled, the optic will<br />

start automatically. A beeping noise will indicate the start of the reaction. Repeat for the<br />

remaining cuvettes. The reaction can also be started either by pressing the optic key or by<br />

using the Autopipette, which electronically triggers the optic channel.<br />

In summary, press the optic key once for patient identification data entry, press<br />

again to activate the channel, and a third time if manually initiating the reaction.<br />

(Do not press a third time if using the Autopipette to start the reaction or if<br />

autostart is active).<br />

<br />

If using the Autopipette,<br />

- always pipette from left to right (channel 1 - 4)!<br />

- Disable the autostart function in the SETUP SYSTEM<br />

29