Operation Manual<br />

<strong>Helena</strong> C-4 - Software C11.11b<br />

<strong>Helena</strong> Biosciences Europe<br />

UK<br />

5.3.10. Result Warning Messages<br />

In addition to the result, the instrument may inform the operator of critical samples by<br />

attaching status characters or error messages.<br />

DISPLAY PRINTER MEANING .<br />

* * Out of calibration (i.e. *167 %)<br />

> | < > Out of scale (i.e. >999.9 mg/dL)<br />

++++ NO CLOT DETECTED No clot detected within 300 seconds<br />

- - - - NO CLOT DETECTED Clot detected before deadtime<br />

. COAGULATION ERROR Detected reaction was not valid for<br />

coagulation<br />

OPTIC LOW SIGNAL Light transmission was not sufficient.<br />

“S”<br />

High sensitivity mode<br />

“R”<br />

Result found<br />

“B”<br />

“F”<br />

Read section on troubleshooting for further details.<br />

Biphasic APTT found (indicative ofDIC)<br />

Low fibrinogen found (indicative of<br />

liver disease)<br />