5.3.10. Result Warning Messages<br />
In addition to the result, the instrument may inform the operator of critical samples by<br />
attaching status characters or error messages.<br />
DISPLAY PRINTER MEANING .<br />
* * Out of calibration (i.e. *167 %)<br />
> | < > Out of scale (i.e. >999.9 mg/dL)<br />
++++ NO CLOT DETECTED No clot detected within 300 seconds<br />
- - - - NO CLOT DETECTED Clot detected before deadtime<br />
. COAGULATION ERROR Detected reaction was not valid for<br />
coagulation<br />
OPTIC LOW SIGNAL Light transmission was not sufficient.<br />
“S”<br />
High sensitivity mode<br />
“R”<br />
Result found<br />
“B”<br />
“F”<br />
Read section on troubleshooting for further details.<br />
Biphasic APTT found (indicative ofDIC)<br />
Low fibrinogen found (indicative of<br />
liver disease)<br />
31