3.4 The calibration system 59<br />

tel-00821629, version 1 - 11 May 2013<br />

Figure 3.14: FADC baseline mean position (left) and RMS (right) as function<br />

of the PMT number. ID PMT are i<strong>de</strong>ntified by PMT number smaller<br />

that 390 while IV PMT by number bigger than 390. This monitoring unit allows<br />

to i<strong>de</strong>ntify abnormal channels through the position and the fluctuation<br />

of the FADC baseline.<br />

3.4.1 Light Injection<br />

LED fibers system is used to inject light of known waveform into the ID an<br />

the IV. The fibers are routed insi<strong>de</strong> the d<strong>et</strong>ector on the edge of some PMTs.<br />

Some injection points are equipped with di↵user p<strong>la</strong>tes to illuminate wi<strong>de</strong>ly<br />

the d<strong>et</strong>ector. The fibers are connected to a blue and UV LEDs f<strong>la</strong>sher which<br />

rate and light intensity is controlled remotely. The light injection system<br />

is used daily to monitor the stability of the PMT gain and their timing<br />

response.<br />

3.4.2 Radioactive Sources<br />

Gamma sources 137 Cs (0.667 MeV), 68 Ge (2⇥0.511 MeV), 60 Co (1.173 and<br />

1.333 MeV) and neutron source 252 Cf, whose rates is at the level of 50 Hz,<br />

have been <strong>de</strong>ployed into the targ<strong>et</strong> and the gamma catcher.<br />

Deployment in the targ<strong>et</strong> is performed along the symm<strong>et</strong>ry axis using a<br />

motorised pulley-and-weight system, operated from a glove-box installed<br />

above the chimney. The <strong>de</strong>ployment range span from 1 cm above the targ<strong>et</strong><br />

bottom up to the targ<strong>et</strong> chimney, with 1 mm precision on the source position.<br />

Deployment in the gamma catcher is performed using a motor driven wire,<br />

gui<strong>de</strong>d through a rigid looped tube. The gui<strong>de</strong> tube traverses the gamma<br />

catcher passing near the boundaries of the targ<strong>et</strong> and the bu↵er. The source<br />

position is known within 1 cm precision along the loop and the perpendicu<strong>la</strong>r<br />

distance b<strong>et</strong>ween the source and the targ<strong>et</strong> wall is known within 2 mm.