Setting Normalized Calcium - iCa(N)<br />
• iCa++ results can be altered by pH.<br />
• Normalizing adjusts iCa++ results for changes in pH.<br />
Most reference labs report iCa(N).<br />
• Air bubble in sample or exposure to room air ↑ pH<br />
and decreases iCa++.<br />
• Delay in processing sample (> 10 min) ↓ pH and<br />
increases iCa++.<br />
• After setting, <strong>IRMA</strong> will report both iCa++ and iCa(N).<br />
• When acid-base status is unknown, care should be<br />
taken when interpreting iCa(N).