Running an LQC Sample (Cont’d)<br />

• Carefully snap open the control ampule<br />

• Draw the control sample into a non-heparinized<br />

syringe using a large bore 18-20 gauge needle.<br />

**Unlike with a patient sample, do not invert the<br />

syringe to expel air bubbles. Keep the needle<br />

pointed down.<br />

• Remove the needle from the syringe and place<br />

syringe in the cartridge port.<br />

• Press test and when prompted gently inject<br />

entire contents of the syringe into the cartridge.<br />

• If QC results are within limits, press done. If<br />

outside the limits, press “notes” and document<br />

problem.