Internal Calibration/Adjustment (Only on SI models)Set the following parameters:Setup: Bal.Scal.: Cal.-Adj.: Cal.Int. (menu code 1.1. 9. 4)The built-in motorized calibration weight is applied and removed automaticallyfor internal calibration.Step Key (or instruction) Display1. Tare the balance t 0.0 g2. Start calibration q Cal.Int.The built-in weight isapplied automaticallyCal. run.3. Calibration/adjustment executed Cal.end4. The built-in weight is removed 0.0 g16