<strong>IHE</strong> Patient Care Device Technical Framework Supplement – Alarm CommunicationManagement (<strong>ACM</strong>)______________________________________________________________________________3.Y+1.4 Interaction Diagrams3.Y+1.4.1 AM status updates to ARARAMACPCD-04 Report AlarmOptPCD-05 Report AlarmStatusPCD-06 Disseminate AlarmOptPCD-05 Report AlarmStatusPCD-07 Report DisseminationAlarm StatusOpt630The AM sends Report Alarm Status to the AR actor as an HL7 ORA message.3.Y+1.4.1.1 Trigger Events635640The AM has determined either through configuration and contextual data driven decision rules orthrough receipt of Dissemination Status from the Alarm Communicator that an alarm statusupdate needs to be sent to the AR.AM internal trigger events include the following:• Accept (not specified, correct)• Reject (not specified, nuisance but correct, false positive)• Deliverable, had a mapped destination• Queued to communications• Undeliverable (AC down or message not deliverable to AC endpoint device)3.Y+1.4.1.2 Message SemanticsThis message is meant to convey from the AM actor to the AR actor the success or failure of theAM and AC to communicate the alarm to the endpoint device.__________________________________________________________________________27Rev. 1.3– 2012-08-16Copyright © 2012: <strong>IHE</strong> International, Inc.