<strong>IHE</strong> Patient Care Device Technical Framework Supplement – Alarm CommunicationManagement (<strong>ACM</strong>)______________________________________________________________________________Figure 3.Z.7.5-1: Event Identification Facet1545155015551560Facet 2. Source identificationFor an event code corresponding with a metric alarm, this segment identifies the particularmeasurement that is the source of the alarm by its MDC nomenclature code in OBX-3Observation Identifier. If it has a numeric value, it shall be in OBX-5 Observation Value, and ifavailable the alarm range set in the device will be encoded in OBX-7 Reference RangFor a technical alarm, this facet specifies the subsystem that is the source of the event by itsMDC object code in OBX-5 Observation Value, and by its dotted sub-ID notation according tothe DEC specification for OBX-4 Observation Sub-ID.Facet 3. Event PhaseEach occurrence contains one of the following phase indications of the alarm from theEventCurrentPhase enumeration:• tpoint• start• continue• end• update• escalate• de-escalate• reset__________________________________________________________________________58Rev. 1.3– 2012-08-16Copyright © 2012: <strong>IHE</strong> International, Inc.