11.07.2015 Views

ACM - IHE

ACM - IHE

ACM - IHE

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>IHE</strong> Patient Care Device Technical Framework Supplement – Alarm CommunicationManagement (<strong>ACM</strong>)______________________________________________________________________________19353.4 <strong>IHE</strong> PCD-06 - wctp-SubmitRequest – no MCRThis message is used to send a message from the <strong>ACM</strong> AM to the <strong>ACM</strong> AC when MCR is not tobe indicated because this is either a test message or the <strong>ACM</strong> AC does not support MCR.1940194519501955notification text3.5 <strong>IHE</strong> PCD-06 - wctp-SubmitRequest – Unpaired MCRThis message is used when the <strong>ACM</strong> AM wants to send a message to the <strong>ACM</strong> AC and whileMCR is to be indicated the <strong>ACM</strong> AC does not support paired MCR so unpaired MCR is used.1960__________________________________________________________________________76Rev. 1.3– 2012-08-16Copyright © 2012: <strong>IHE</strong> International, Inc.