11.07.2015 Views

ACM - IHE

ACM - IHE

ACM - IHE

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>IHE</strong> Patient Care Device Technical Framework Supplement – Alarm CommunicationManagement (<strong>ACM</strong>)______________________________________________________________________________1910The assumption to this response is that the <strong>ACM</strong> AM is to use only WCTP 1.1 XML messagesand not later, e.g. is no support for MCR.3.3 Administrative - wctp-VersionResponse1915This message is used when Version Query operation is supported.192019251930A response dtdName of "wctp-dtd-v1r3" indicates support for <strong>ACM</strong> profile conformant WCTPversion 1.3 which indicates support for Multiple Choice Response (MCR) pairs on WCTPSubmitRequest messages. MCR pairs are used to populate soft keys on wireless device operatorinterfaces and so that the reply value can be vendor specific and still be presented in a vendoragnostic manner. A response of dtdName of "wctp-dtd-v1r2" indicates support for WCTP 1.2which supports non-paired MCR.__________________________________________________________________________75Rev. 1.3– 2012-08-16Copyright © 2012: <strong>IHE</strong> International, Inc.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!