I/O <strong>Module</strong> Introduction6. CJC Calibration (only for T/C input module)Figure 25: Cold junction calibration* Note: Zero calibration and span calibration must be completedbefore CJC calibration. To calibrate CJC, the thermocoupleattached to <strong>ADAM</strong>-5018 and a standard thermometershould be used to measure a standard known temperature,such as the freezing point of pure water. The amount ofoffset between the <strong>ADAM</strong>-5018 and the standardthermometer is then used in the <strong>ADAM</strong> utility to completeCJC calibration.P.32 <strong>ADAM</strong>-<strong>5000</strong> <strong>Series</strong> I/O <strong>Module</strong> User’s <strong>Manual</strong>