Operations 47. If the O 2cell current is too large to be zeroed, an errormessage will appear. Press ACK to retry or ESC toabort the zeroing process.8. If AUTO was chosen in step 4, wait for the O 2level tosettle before the system zeros. This may take from thirty(30) seconds to fifteen (15) minutes, depending on thesample gas <strong>oxygen</strong> level.9. After a stable low <strong>oxygen</strong> level is detected by the system,the system will automatically zero.If the O 2cell current is too large to be zeroed, an errormessage will appear (see step 7). You will then have topress ESC to abort, or ACK to retry.10. If the zero is successful, the system will return to themain Analyzer mode menu.TELEDYNE ANALYTICAL INSTRUMENTS4-7