Principles and Practical Aspects of Preparative Liquid Chromatography
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
volume <strong>of</strong> both systems. The linear method transfer from an analytical to<br />
the preparative system was accomplished using the formulas in Section<br />
4.2 “Formulas for linear scale-up from analytical to preparative columns”.<br />
Equilibration Isocratic steps Gradient Purge<br />
100%B<br />
t iso1<br />
0%B<br />
Time<br />
Figure 4.1 Gradient pr<strong>of</strong>ile for an analytical scouting run using a 4.6-mm id column,<br />
comprising injection, optional isocratic hold, gradient slope <strong>and</strong> purge phases.<br />
Equilibration<br />
Isocratic steps<br />
Gradient<br />
Purge<br />
100%B<br />
t h2<br />
t iso2<br />
0%B<br />
Time<br />
Figure 4.2 Gradient pr<strong>of</strong>ile after scale-up to a 21.2-mm id column, comprising<br />
an extended isocratic step to compensate for the applied flow rate with respect to<br />
the difference in dwell volumes between the analytical <strong>and</strong> preparative system.<br />
52