EasyOne Pro LAB only<br />

Danger<br />

Explosion Hazard—<br />

Introduction<br />

� Ensure that persons working with oxygen have undergone<br />

special training and are aware of the particular properties<br />

of oxygen to guarantee safe handling of oxygen without<br />

accidents. Furthermore, these persons must be familiar<br />

with this operator manual and must have understood its<br />

content.<br />

� Do not use hand cream or other skin care products.<br />

� Oxygen intensively promotes combustion; therefore,<br />

flammable substances must be kept away from oxygen.<br />

� Unintended release of oxygen in confined spaces<br />

increases risk of fire; smoking and open flames are<br />

prohibited. Possible sources of ignition must be<br />

eliminated. Rooms must be well ventilated.<br />

� The oxygen inlet must periodically be checked for leaks<br />

(at least once a month).<br />

� Use only fittings approved for use with oxygen; these<br />

fittings must be properly connected. All fittings,<br />

downstream lines and devices must be free of oil and<br />

grease.<br />

� Connect a pressure relief valve to the pressurized gas<br />

cylinder before opening it! Before connecting the<br />

pressure relief valve, check the connection sleeve of the<br />

pressurized gas cylinder for contamination. Contaminated<br />

connections must be cleaned with a clean cloth. The valve<br />

of the pressurized gas cylinder may be opened only if the<br />

connected fittings are closed. Open the valve of the<br />

compressed gas cylinder very slowly.<br />

12 EasyOne Pro / EasyOne Pro LAB