EasyOne Pro LAB only<br />
Danger<br />
Explosion Hazard—<br />
Introduction<br />
� Ensure that persons working with oxygen have undergone<br />
special training and are aware of the particular properties<br />
of oxygen to guarantee safe handling of oxygen without<br />
accidents. Furthermore, these persons must be familiar<br />
with this operator manual and must have understood its<br />
content.<br />
� Do not use hand cream or other skin care products.<br />
� Oxygen intensively promotes combustion; therefore,<br />
flammable substances must be kept away from oxygen.<br />
� Unintended release of oxygen in confined spaces<br />
increases risk of fire; smoking and open flames are<br />
prohibited. Possible sources of ignition must be<br />
eliminated. Rooms must be well ventilated.<br />
� The oxygen inlet must periodically be checked for leaks<br />
(at least once a month).<br />
� Use only fittings approved for use with oxygen; these<br />
fittings must be properly connected. All fittings,<br />
downstream lines and devices must be free of oil and<br />
grease.<br />
� Connect a pressure relief valve to the pressurized gas<br />
cylinder before opening it! Before connecting the<br />
pressure relief valve, check the connection sleeve of the<br />
pressurized gas cylinder for contamination. Contaminated<br />
connections must be cleaned with a clean cloth. The valve<br />
of the pressurized gas cylinder may be opened only if the<br />
connected fittings are closed. Open the valve of the<br />
compressed gas cylinder very slowly.<br />
12 EasyOne Pro / EasyOne Pro LAB