spirette<br />

Start-up and Initial Preparation<br />

Note<br />

The device is to be operated only with the original <strong>ndd</strong><br />

spirette.<br />

Warning<br />

Cross Contamination Between Patients—The spirette is<br />

intended for single patient use. Use a new one for each<br />

patient to prevent cross contamination between patients.<br />

� Tear open the plastic bag containing the spirette and fold the bag back<br />

so that you can insert the spirette in the flow sensor. Ensure that the<br />

plastic bag protects the mouthpiece of the spirette until you hand the<br />

flow sensor over to the patient.<br />

This approach not only ensures perfectly hygienic conditions, it also keeps<br />

the spirette closed for subsequent setting of the baseline.<br />

� Insert the spirette - as shown in the illustration - into the flow sensor<br />

as far as possible. When doing this, please ensure that the arrow on the<br />

spirette is lined up with the arrow on the flow sensor.<br />

� You remove the spirette by pushing it out of the sensor from below.<br />

Fig. 2-22 Inserting the spirette<br />

36 EasyOne Pro / EasyOne Pro LAB