Start-up and Initial Preparation<br />
FRC Gas (100% O2), EasyOne Pro LAB only<br />
<strong>Medical</strong> grade oxygen is used for FRC measurements with EasyOne Pro<br />
LAB.<br />
Secondary pressure: 3 bar<br />
Flow: > 0.7 l/s (at 3 bar)<br />
Warning<br />
Improper handling of gas cylinders represents a major risk to<br />
persons and the environment. The relevant regulations must<br />
be observed without exception. Protect the gas cylinders<br />
from falling. The content and fill level must be clearly visible<br />
at all times. Valves, fittings, connections and tubing must be<br />
free of oil and grease. Have the cylinder inspected regularly<br />
by an official test authority. The user is liable for any<br />
damage. Close the main valve whenever the gas cylinder is<br />
not in use!<br />
� Check that the main valve a and the pressure relief valve c are closed.<br />
Fig. 2-12 Gas cylinder fittings<br />
a Gas cylinder main valve<br />
b Manometer, primary side<br />
c Pressure relief valve, secondary side<br />
a b c d e<br />
28 EasyOne Pro / EasyOne Pro LAB<br />
d Manometer, secondary side<br />
e Port for EasyOne Pro LAB connection<br />
tube (FRC gas)