Start-up and Initial Preparation<br />

FRC Gas (100% O2), EasyOne Pro LAB only<br />

<strong>Medical</strong> grade oxygen is used for FRC measurements with EasyOne Pro<br />

LAB.<br />

Secondary pressure: 3 bar<br />

Flow: > 0.7 l/s (at 3 bar)<br />

Warning<br />

Improper handling of gas cylinders represents a major risk to<br />

persons and the environment. The relevant regulations must<br />

be observed without exception. Protect the gas cylinders<br />

from falling. The content and fill level must be clearly visible<br />

at all times. Valves, fittings, connections and tubing must be<br />

free of oil and grease. Have the cylinder inspected regularly<br />

by an official test authority. The user is liable for any<br />

damage. Close the main valve whenever the gas cylinder is<br />

not in use!<br />

� Check that the main valve a and the pressure relief valve c are closed.<br />

Fig. 2-12 Gas cylinder fittings<br />

a Gas cylinder main valve<br />

b Manometer, primary side<br />

c Pressure relief valve, secondary side<br />

a b c d e<br />

28 EasyOne Pro / EasyOne Pro LAB<br />

d Manometer, secondary side<br />

e Port for EasyOne Pro LAB connection<br />

tube (FRC gas)