CAMLOG ® IMPLANT SYSTEM<br />
CAMLOG ® PROSTHETICS<br />
TORQUE WRENCH SETTINGS<br />
FOR FINAL FIXATION OF SURGICAL COMPONENTS AND SUPRASTRUCTURES<br />
Article<br />
CAMLOG ® Implant<br />
cover screw:<br />
CAMLOG ® Healing caps (incl. PS):<br />
cylindrical<br />
wide body<br />
bottleneck<br />
Vario SR protection caps<br />
CAMLOG ® Lab screw:<br />
CAMLOG ® Vario SR Lab screw:<br />
CAMLOG ® Temporary abutment<br />
(incl. PS)<br />
CAMLOG ® Scanbody<br />
CAMLOG ® Abutment screw:<br />
CAMLOG ® Vario SR Abutment screw:<br />
CAMLOG ® Esthomic ® abutment,<br />
straight (incl. PS)<br />
CAMLOG ® Esthomic ® abutment,<br />
angled 15°/20° (incl. PS)<br />
CAMLOG ® Esthomic ® abutment, Inset<br />
CAMLOG ® Universal abutment (incl. PS)<br />
CAMLOG ® Gold-plastic abutment<br />
CAMLOG ® Titanium Bases CAD/CAM<br />
CAMLOG ® Ceramic abutment<br />
CAMLOG ® Logfit ® abutment<br />
CAMLOG ® Telescope abutment<br />
(from left to right)<br />
CAMLOG ® Vario SR abutment, straight<br />
CAMLOG ® Vario SR abutment,<br />
20°/30° angled<br />
All screws have to be tightend after at least 5 minutes with the corresponding torque!<br />
52<br />
J5320.1030<br />
tightened by hand<br />
20 Ncm<br />
Vario SR prosthetic screw: (Ø 3.8/4.3/5.0/6.0 mm)<br />
15 Ncm<br />
Vario SR<br />
Plastic copings/titanium caps/<br />
bases for bar<br />
J4006.1601<br />
(Ø 3.3/3.8/4.3 mm)<br />
J4008.1600<br />
(Ø 3.8/4.3 mm)<br />
J4005.1601<br />
(Ø 3.3/3.8/4.3 mm)<br />
J4007.1600<br />
(Ø 3.8/4.3 mm)<br />
J4005.2004<br />
J4006.2001<br />
(Ø 5.0/6.0 mm)<br />
J4008.2000<br />
(Ø 5.0/6.0 mm)<br />
J4005.2001<br />
(Ø 5.0/6.0 mm)<br />
J4007.2000<br />
(Ø 5.0/6.0 mm)<br />
J5316.0510 (hex, extra short)<br />
J5316.0501 (hex, short)<br />
J5316.0502 (hex, long)