<strong>Manual</strong> <strong>do</strong> Usuário LEGENDAIR ®<<strong>br</strong> />
Corte da AUSÊNCIA REDE<<strong>br</strong> />
alimentação<<strong>br</strong> />
elétrica AUSÊNCIA ALIM.<<strong>br</strong> />
ELÉTRICA<<strong>br</strong> />
Bateria interna<<strong>br</strong> />
Medida de<<strong>br</strong> />
pressão<<strong>br</strong> />
Válvula de<<strong>br</strong> />
expiração<<strong>br</strong> />
Memória<<strong>br</strong> />
interna<<strong>br</strong> />
Tecla<strong>do</strong><<strong>br</strong> />
Parada<<strong>br</strong> />
involuntária<<strong>br</strong> />
BATERIA<<strong>br</strong> />
DECARREGADA<<strong>br</strong> />
FIM BATERIA<<strong>br</strong> />
VERIFICAR<<strong>br</strong> />
BATERIA<<strong>br</strong> />
Corte da alimentação da rede AC<<strong>br</strong> />
Comutação à alimentação externa DC ou bateria interna.<<strong>br</strong> />
Corte da alimentação externa DC<<strong>br</strong> />
Comutação à bateria interna.<<strong>br</strong> />
Nível de carga da bateria interna < 5% Ou Tensão de<<strong>br</strong> />
alimentação < 23,8 V se bateria <strong>com</strong> defeito<<strong>br</strong> />
Nível de carga da bateria interna = 0% Ou Tensão de<<strong>br</strong> />
alimentação < 21,8 V se bateria <strong>com</strong> defeito<<strong>br</strong> />
Parada imediata da ventilação<<strong>br</strong> />
Impossível carregar a bateria interna<<strong>br</strong> />
Anomalia de controle das tensões de alimentação durante mais<<strong>br</strong> />
de 1 min 30 s<<strong>br</strong> />
Pressão anormal: Pressão constante ou negativa durante mais<<strong>br</strong> />
VERIFICAR<<strong>br</strong> />
de 45 s (salvo em caso de desligamento ativo)<<strong>br</strong> />
PRESSÃO Pressão anormal: Pressão constante ou negativa durante mais<<strong>br</strong> />
de 2 min (salvo em caso de desligamento ativo)<<strong>br</strong> />
Pressão interna válvula anormal durante 15 s<<strong>br</strong> />
VERIFICAR<<strong>br</strong> />
Ou Pep medida > Pe+ 10 mbar durante 45 s<<strong>br</strong> />
VÁLVULA Pressão interna válvula anormal<<strong>br</strong> />
Ou Pep medida > Pe + 10 mbar durante 2 min<<strong>br</strong> />
FUGA VÁLVULA<<strong>br</strong> />
VERIFICAR<<strong>br</strong> />
AJUSTES<<strong>br</strong> />
VERIFICAR<<strong>br</strong> />
TECLAS<<strong>br</strong> />
-<<strong>br</strong> />
Débito expira<strong>do</strong> anormalmente eleva<strong>do</strong> durante a fase de<<strong>br</strong> />
inspiração (em versão tubo duplo)<<strong>br</strong> />
Perda de parâmetros memoriza<strong>do</strong>s<<strong>br</strong> />
Este alarme é sistemático após o carregamento de um novo<<strong>br</strong> />
software.<<strong>br</strong> />
I - 77 – 00<<strong>br</strong> />
MÉDIA SIM<<strong>br</strong> />
MÉDIA SIM<<strong>br</strong> />
ALTA SIM<<strong>br</strong> />
ALTA NÃO<<strong>br</strong> />
MÉDIA SIM<<strong>br</strong> />
MÉDIA SIM<<strong>br</strong> />
ALTA SIM<<strong>br</strong> />
MÉDIA SIM<<strong>br</strong> />
ALTA SIM<<strong>br</strong> />
MÉDIA SIM<<strong>br</strong> />
MÉDIA SIM<<strong>br</strong> />
Tecla <strong>do</strong> tecla<strong>do</strong> mantida premida por mais de 20 s (*) ALTA NÃO<<strong>br</strong> />
Corte da alimentação elétrica pelo interruptor principal quan<strong>do</strong> a<<strong>br</strong> />
ventilação está em curso<<strong>br</strong> />
ALTÍS-<<strong>br</strong> />
SIMA<<strong>br</strong> />
Nota: Em caso de corte da alimentação elétrica pelo interruptor quan<strong>do</strong> a ventilação<<strong>br</strong> />
está em curso, o aparelho volta a arrancar diretamente em ventilação ao ser<<strong>br</strong> />
restabelecida a alimentação.<<strong>br</strong> />
ATENÇÃO<<strong>br</strong> />
(*) Em caso de bloqueio das teclas ou , o início efetivo de to<strong>do</strong>s os<<strong>br</strong> />
alarmes é temporiza<strong>do</strong> de 20 segun<strong>do</strong>s.<<strong>br</strong> />
ANOMALIAS TÉCNICAS<<strong>br</strong> />
As anomalias técnicas não afetam diretamente ou a curto prazo o funcionamento da<<strong>br</strong> />
máquina e não são sinalizadas diretamente ao utiliza<strong>do</strong>r. A consulta das anomalias<<strong>br</strong> />
técnicas é efetuada através <strong>do</strong> menu de Manutenção (ver § Manutenção).<<strong>br</strong> />
A cada tipo de anomalia corresponde um número. Estas informações poderão ser<<strong>br</strong> />
utilizadas pelos técnicos especializa<strong>do</strong>s Quan<strong>do</strong> das inspeções periódicas.<<strong>br</strong> />
NÃO