2. Empty the measuring cell by pressing the pump lever.3. Open the pump lock and remove the pump.4. Hold the pump housing under running water and rinse it thoroughly.5. Insert the pump and close the pump lock.6. Empty the measuring cell by pressing the pump lever.EN10.4 Cleaning the Housing and DisplayTo clean the housing and display, use a soft cloth dipped in ethanol or warm water. Ifnecessary, you can use a mild solvent (pH < 10).10.5 Storing the InstrumentBefore storing the instrument for a longer period of time, clean the measuring cell asdescribed in Chapter 10.1. This prevents any liquid residues freezing or drying in themeasuring cell, which could destroy the cell or lead to measuring inaccuracies.For storage lasting less than one day, fill the measuring cell with deionized water orsolvent. If you have filled the liquid via the plastic syringe, leave the syringe in theadapter to stop the liquid running out.C96IB01 55