2.5 Electronic Control<br />

The electronics controlled by two microprocessors allows extensive adaptations<br />

of the centrifuge to the different tasks. The following parameters can be<br />

programmed and displayed among others:<br />

− Speed (by activation of FINE steps of 1 or 10 rpm possible)<br />

− RCF in steps of 1 or 10 x g<br />

− Time preselection (9 h, 59 min max.), in steps of 1 min or 1 sec<br />

− Continuous operation<br />

− Short-time operation<br />

− Fixed deceleration and acceleration curves<br />

− Free creation of deceleration and acceleration curves<br />

− Saving, recalling and alteration of programs<br />

− Input and measurement of the time integral<br />

− Start delay<br />

− Continuous self-monitoring and recognition of errors which are displayed and<br />

saved for service<br />

2.6 Safety Devices<br />

Apart from the passive safety devices due to the instrument's mechanical design<br />

there are the following active precautions for your safety:<br />

2.6.1 Lid Lock, Cover Closing Device<br />

The centrifuge can only be started when the power switch is switched to "ON“<br />

and when the lid is correctly closed. After closing the lid the motorized lid lock is<br />

automatically locked. The lid can only be opened when the rotor has completely<br />

stopped. If the lid is opened by the emergency release during operation, the<br />

centrifuge will immediately switch off and decelerate brakeless up to standstill of<br />

the rotor. If the lid is open, the drive is completely separated from the mains<br />

supply, that means starting of the centrifuge is impossible (refer to chapter 9.5.4<br />

"Emergency lid release").<br />

2.6.2 Imbalance Monitoring System<br />

In the event that uneven loading leads to imbalance, the drive is switched off and<br />

an imbalance warning message is displayed.<br />