4.2.4 Time<br />
Time<br />
The set run time is displayed in the upper section of this area, underneath the<br />
remaining run time is displayed. Time is defined as the period from the start of<br />
the centrifuge to the beginning of deceleration, maximum value is 9 h 59 min.<br />
The set time is underlined (here: 2 minutes).<br />
The set value is indicated in hours, minutes, and seconds. The actual value has<br />
the same units as the set value and is displayed in hours : minutes or in minutes :<br />
seconds if the set value is below 10 minutes.<br />
h:m<br />
m:s<br />
If the highest possible time of 9 h 59 min is exceeded or below the minium<br />
adjustable time range, continuous operation is activated. The word "HOLD" is<br />
displayed instead of the set value. After the start of a continuous run, the elapsed<br />
time is displayed instead of the remaining run time. By entering a specific run<br />
time the continuous mode is deactivated. It can be terminated by pressing the<br />
Stop-key as well.<br />
Continuous run<br />
<strong>Operating</strong> <strong>Manual</strong> SIGMA 3-<strong>18</strong>, page 47 of 95<br />
03/06