<strong>UltiMate</strong> <strong>3000</strong> Series:<br />
<strong>SRD</strong>-<strong>3x00</strong> <strong>Solvent</strong> <strong>Racks</strong><br />
2.6 Leak Sensor<br />
A leak sensor is installed inside the <strong>SRD</strong>. If liquid collects in the drip tray under the fluid<br />
connections, the leak sensor reports a leak, and the Status LED on the front panel door<br />
changes to red.<br />
If the <strong>SRD</strong> is operated from a pump of the <strong>UltiMate</strong> <strong>3000</strong> system and if the leak sensor is<br />
exposed to liquid for at least 30 seconds, a message (<strong>Solvent</strong> Rack leak detected) appears on<br />
the front panel display of the pump and a beep sounds. If the pump is operated with<br />
Chromeleon, a message is also displayed in the Chromeleon Audit Trail.<br />
When the leak sensor reports a leak, eliminate the cause for the leakage and dry the leak<br />
sensor (→ page 43).<br />
2.7 Wellness<br />
System Wellness monitors the health of the <strong>SRD</strong>. Therefore, the <strong>SRD</strong> supports several<br />
performance and reliability features that can help you detect small problems before they<br />
turn into big ones:<br />
• Leak detection–Status LED on the front panel door<br />
• Degasser vacuum level monitoring–Status and Vacuum LEDs on the front panel door<br />
When an error is detected, the Status LED or the Status and Vacuum LEDs on the front<br />
panel door change to red. If the <strong>SRD</strong> is operated from the pump of the <strong>UltiMate</strong> <strong>3000</strong> system,<br />
one or more messages appear on the pump display and in the Chromeleon Audit Trail<br />
(→ page 35).<br />