<strong>UltiMate</strong> <strong>3000</strong> Series:<br />

<strong>SRD</strong>-<strong>3x00</strong> <strong>Solvent</strong> <strong>Racks</strong><br />

4.3 <strong>Solvent</strong> Reservoirs<br />

All <strong>Solvent</strong> <strong>Racks</strong> are shipped with solvent reservoirs and appropriate tubing, including frit<br />

holders with filter frits.<br />

4.3.1 General Notes<br />

When connecting the solvent reservoirs, observe the following general precautions:<br />

• Observe the precautions for solvent selection (→ page 25).<br />

• Before using the solvent reservoirs for the first time, rinse them thoroughly by using<br />

high-purity solvents.<br />

• Always install filter frits on the solvent supply line filters. This prevents contaminants<br />

from reaching the HPLC system that may increase wear and cause damage to the system.<br />

• Regularly check the filter frits for permeability. This is especially important when using<br />

aqueous solvents. Aqueous solvents may contaminate the filters with algae and other<br />

microorganisms that deposit on the filter frits. Therefore, use fresh solvents at regular<br />

intervals. Clean the reservoirs thoroughly before refilling them and replace the filter frits<br />

as necessary.<br />

• As a standard, filter holders with stainless steel frits are provided in the accessories kit of<br />

the <strong>Solvent</strong> Rack. Do not use these frits if the system includes a biocompatible pump, and<br />

NCS-3500RS or an NCP-3200RS. In this case, replace the stainless steel frits with the<br />

frits from the accessories kit of the module.<br />

Open the filter holder and remove the filter frit. When placing the new filter frit into the<br />

bottom part, make sure that the frit is in a level position (avoid tilting the frit).<br />

• Make sure that the tubing connecting the pump to the degasser is as short as possible and<br />

locate the solvent reservoirs as close as possible to the pump. To avoid formation of air<br />

bubbles in the reservoirs and reformation of air bubbles in the solvent, make sure that the<br />

reservoirs are on the same level or higher as the pump. Therefore, stack the rack onto the<br />

pump as shown in Fig. 4 (→ page 16).<br />

• Normal phase eluents usually show only a low concentration of dissolved gases.<br />

Therefore, it is normally not required to use a degasser with these eluents.<br />

• Before connecting the solvent supply lines, make sure that the connectors are free of<br />

contaminants. Even minute particles can allow air to enter the degasser, and thus reduce<br />

the degassing effectiveness.<br />