Problem Probable Cause Remedial Action<br />

Poor peak area precision<br />

(Cont'd)<br />

The environmental conditions are<br />

unstable.<br />

Contamination occurs somewhere<br />

in the system.<br />

For additional causes, refer to the<br />

operating instructions for the<br />

autosampler and for the pump.<br />

Poor degassing There is a leak in the capillaries or<br />

solvent supply lines.<br />

Make sure that the temperature<br />

and the humidity are constant.<br />

Consider using a thermostatted<br />

column compartment. Avoid<br />

draft.<br />

Flush the system using an<br />

appropriate solvent. Check the<br />

filter frits in the solvent supply<br />

line filters for contamination.<br />

Replace the frits as necessary.<br />

→ Autosampler manual<br />

→ Pump manual<br />

→ NCS/NCP manual<br />

Inspect the capillaries and<br />

solvent supply lines for<br />

leakage; tighten loose fitting<br />

connections.<br />

The flow rate is too high. Reduce the flow rate.<br />

Degasser noise The vacuum pump of the degasser<br />

is running at high speed.<br />

Fill and degas all channels<br />

(even if they are not used for<br />

the application).<br />

Page 40 Operating Instructions