<strong>UltiMate</strong> <strong>3000</strong> Series:<br />

<strong>SRD</strong>-<strong>3x00</strong> <strong>Solvent</strong> <strong>Racks</strong><br />

Problem Probable Cause Remedial Action<br />

High baseline drift<br />

(Cont'd)<br />

High noise level, nonperiodic<br />

baseline<br />

fluctuation<br />

Periodic baseline<br />

fluctuation, pulsation<br />

Reproducible ghost peaks<br />

in the chromatogram.<br />

Additional peaks appear<br />

in the injection peak.<br />

The eluents are degraded or<br />

inhomogeneous.<br />

The environmental conditions are<br />

unstable.<br />

The mobile phase is delivered in<br />

circles.<br />

For additional causes, refer to the<br />

operating instructions for the<br />

detector.<br />

The eluent is degraded or of poor<br />

quality/purity.<br />

For additional causes, refer to the<br />

operating instructions for the<br />

detector.<br />

There are pressure fluctuations<br />

from the pump.<br />

Before you start an analysis, be<br />

sure that the eluents are already<br />

homogenized in the reservoirs.<br />

Use fresh solvent and check the<br />

eluent filter frits. In aqueous<br />

solvents, growth of<br />

microorganisms is possible.<br />

Make sure that the temperature<br />

and the humidity are constant.<br />

Avoid draft. Verify on the<br />

detector that the lamp and flow<br />

cell covers are in proper<br />

position and that the front panel<br />

door is closed.<br />

Direct the mobile phase to<br />

waste.<br />

→ Detector manual<br />

Use fresh solvents. Use HPLC<br />

grade solvents (MS grade if the<br />

<strong>SRD</strong> is operated together with<br />

an NCS-3500RS or<br />

NCP-3200RS).<br />

→ Detector manual<br />

Purge the pump<br />

(→ Pump manual).<br />

There are air bubbles in the system. Purge the pump<br />

(→ Pump manual).<br />

The degassing channels are<br />

contaminated.<br />

The eluents are degraded, dirty or<br />

of poor purity/quality.<br />

Contamination occurs somewhere<br />

in the system.<br />

With gradients, the equilibration<br />

time after the flush cycle is too<br />

short.<br />

Rinse the degassing channels<br />

(→ page 42).<br />

Use fresh eluent. Use HPLC<br />

grade solvents (MS grade if the<br />

<strong>SRD</strong> is operated together with<br />

an NCS-3500RS or<br />

NCP-3200RS).<br />

Flush the system using an<br />

appropriate solvent.<br />

Extend the equilibration time.<br />

There is excessive dead volume. Eliminate possibly existing<br />

dead volume.<br />

Page 38 Operating Instructions