3.4 Connecting the <strong>SRD</strong><br />
3.4.1 Connecting the <strong>SRD</strong> to an <strong>UltiMate</strong> <strong>3000</strong> System Pump<br />
Locate the connection cable in the accessories kit of the <strong>SRD</strong> and connect the 15-pin D-<br />
Sub port on the rear panel with the <strong>Solvent</strong> Rack port on the rear panel of the pump. To<br />
ensure trouble-free operation, use only the cable shipped with the instrument.<br />
When the <strong>SRD</strong> is connected to the pump, the solvent rack is powered from the pump. The<br />
<strong>SRD</strong> need not be connected directly to the main power supply.<br />
For information about the pin assignment for the 15-pin D-Sub connector, see page 51.<br />
3.4.2 Connecting the <strong>SRD</strong> to the Main Power Supply (Optional)<br />
The <strong>SRD</strong> needs to be connected to the main power supply only if it is not operated from an<br />
<strong>UltiMate</strong> <strong>3000</strong> system pump. For this scenario, an external power supply (part no.<br />
6510.0004) is available as an option for the <strong>SRD</strong>.<br />
Use the external power supply unit to connect the DC input on the rear panel to a power<br />
source that is connected to a true ground. No manual adjustment is required to adapt the line<br />
voltage to local voltage requirements.<br />
Important: Do not connect the external power supply unit when the <strong>SRD</strong> is<br />
connected to an <strong>UltiMate</strong> <strong>3000</strong> system pump (→ section 3.4.1).<br />
Important: N'utilisez pas d'alimentation électrique externe si l'instrument est<br />
raccordé un pompe du système <strong>UltiMate</strong> <strong>3000</strong> (→ chapitre 3.4.1).<br />