Processing Product:<br />

a. Note: always close the Buffer and Filtrate valve before starting or stopping the <strong>PureTec</strong>.<br />

Fill the Diafiltration Vessel with the solution to be processed and close the container.<br />

b. If needed, refill the Exchange Buffer Reservoir and empty the Filtrate Reservoir. Place<br />

the tubing back in both containers.<br />

c. If required, edit the Pump Rate and other parameters entered during the priming process.<br />

Press “Execute” and then “RUN” from the front panel to start the pump. If the system<br />

was not previously primed, a decrease in volume in the Vessel will be observed. This<br />

volume will be recovered at the end of the process.<br />

d. To concentrate the product, keep the Buffer Valve closed and open the Filtrate Valve<br />

and Air Vent Valve. This will reduce the volume in the Diafiltration Vessel as the Filtrate<br />

is removed, thus concentrating the solution. When the desired concentration volume has<br />

been achieved, close the Filtrate and Air Vent Valves.<br />

e. To wash the concentrated solution, open only the Filtrate and Buffer Valves. Once the<br />

solution is concentrated, the <strong>PureTec</strong> could be stopped, the Cumulative Counters<br />

cleared, and the Filtrate Weight alarm set to stop the pump when the desired amount of<br />

Exchange Buffer has been collected in the Filtrate Reservoir. The balance also allows<br />

you to monitor the Permeate Flow Rate (FP), which can be optimized by altering the<br />

pump rate on the fly from the front panel. Press the “RATE” key, increase or decrease<br />

the rate as needed, and press “C” to select it. The Exchange Buffer will be added at the<br />

same rate the Filtrate is removed by the vacuum created in the Diafiltration Vessel.<br />

If there is more solution to process than the Vessel would originally hold, more can be<br />

added by connecting the Exchange Buffer tubing to a container with additional process<br />

solution. When this container is empty, the pump could be stopped, the tubing<br />

reconnected to the Exchange Buffer, and the solution either concentrated further or<br />

washed.<br />

f. To empty the Diafiltration Vessel at the end of the run, close the Buffer and Filtrate<br />

valves, and Stop the pump. Close the Valve on the Retentate return line at the bottom of<br />

the Vessel. Have a container ready for the transfer. Disconnect the return line at the<br />

valve, and carefully place it in the container. Open the Air Vent valve, and press RUN on<br />

the pump, this will pump the contents of the Diafiltration Vessel out through the pump and<br />

filter cartridge and into the transfer container. Re-connect the tubing, open the Retentate<br />

valve again, and your system is ready for cleaning.<br />

Documentation:<br />

The <strong>PureTec</strong> displays and prints out several filtration-related parameters. The user can scroll<br />

though three display options providing an instantaneous overview of the filtration progress<br />

and status. All information is printed out in an “Excel-friendly” format. Alternatively, our<br />

SciDoc Software, which uses WinWedge software in conjunction with a customized Excel<br />

spreadsheet, can be used to summarize the filtration data and create graphical<br />

representations of the data on your PC as shown below:<br />