6.0 TANDEM Dual Channel Pump Head:<br />
The TANDEM peristaltic pump head is specifically designed for use with the<br />
<strong>PureTec</strong> system. The SciLog TANDEM pump head will provide you with rugged<br />
reliability as long as common sense maintenance and good quality pump tubing are<br />
used. For filtration applications, you should be using either Platinum-Cured Silicone<br />
or PharMed pump tubing in the correct sizes.<br />
The TANDEM pump head is driven by an 8, 160, or 600-RPM, high-torque motor.<br />
The pump motor is optically encoded and servo-controlled, thus the TANDEM pump<br />
head will maintain a constant output over a wide range of filtration conditions.<br />
However, when the pump head requires excessive torque because of pump tube<br />
failure or “freezing” of the pump head, then the <strong>PureTec</strong> control software will<br />
recognize this condition and go into a stand-by mode, the pump motor is turned off<br />
and the following message is displayed:<br />
CHECK PUMP HEAD<br />
Press Any Key<br />
Before continuing with your pumping application, remove the defective pump head /<br />
tubing and either clean or replace with a functional pump head. This feature<br />
(PumpSense) has been implemented by SciLog to protect your pump motor and<br />
electronics. NOTE: There is nothing wrong with the <strong>PureTec</strong> controller when<br />
you see this display. The problem lies with the pump head and /or pump<br />
tubing you are using.<br />
When you “Press Any Key” to leave the stand-by mode, you will enter the Main<br />
Menu. After you check and replaced your pump head / tubing, you may re-initialize<br />
your application.<br />