2.3 RATE: Constant Rate Filtration; Execute Display<br />
Mode Select RATE<br />
Up Down Select<br />
Select<br />
- Const. RATE Mode -<br />
Exec Edit Prime<br />
Exec<br />
SCALE INITIALIZATION<br />
Please Wait<br />
SET: T14 PR: 100 ml/m<br />
Press RUN when Ready<br />
"RUN"<br />
REMOVING TARE WEIGHT<br />
Please Wait<br />
09:30:15 P1 0.0 RUN<br />
CV 0.0 FF 100.0<br />
The "Display/Switch" key on the front panel<br />
allows you to change between 3 operational<br />
display screens.<br />
Abbreviations are as follows:<br />
P1= Inlet Line Pressure<br />
P2= Retentate Line Pressure<br />
P3= Filtrate Line Pressure<br />
TM= Trans-Membrane Pressure<br />
CV= Cumulative Volume<br />
FF= Flow Rate, Feed<br />
FQ= Filtrate Quantity<br />
FP= Flow Rate, Permeate (Collection Rate)<br />
FR= Flow Rate, Recirculation<br />
T = Temperature (By in-line probe)<br />
A1 = Conductivity from SciCon<br />
A2 or A3 = Temperature from SciCon<br />
RUN= Pump Status<br />
Unless a Balance is connected and properly<br />
configured, and the Filtrate Weight Alarm is<br />
enabled, the Filtrate Quantity and Flow Rate,<br />
Permeate displays will be zero. The "Scale<br />
Initialization" and "Removing Tare Weight"<br />
screens, will only occur in this case as well.<br />
"SWITCH" "STOP"<br />
P1 0.0 P2 0.0 09:31:15 P1 20.0 STOP<br />
P3 0.0 TM 0.0 CV 100.0 FF 100.0<br />
"SWITCH"<br />
T 24.3 FP 0.0 PUMPING STOPPED<br />
FQ 0.0 FR 0.0 Press RUN to Finish<br />
"SWITCH" "EXIT"<br />
A1 0.0 A2 0.0 - Const. RATE Mode -<br />
A3 0.0 Exec Edit Prime<br />
41