5.3 Run Screen (Cont’d)<br />

6870 Freezer/Mill<br />

During the PRECOOL stage the TIME REMAINING line counts down the time for that stage in<br />

1-second increments.<br />

After the PRECOOL stage comes the first GRINDING stage, and the screen displays STATUS:<br />

GRINDING. The TIME REMAINING line will display the time left in that grinding stage. If there<br />

are 3 grinding periods, during the first one the CYCLES REMAINING line will display 3 OF 3,<br />

during the second the CYCLES REMAINING line will display 2 of 3, etc. In every stage the TIME<br />

REMAINING line will count down the time left in that stage in 1-second increments.<br />

If the program on the CURRENT SETTINGS screen shown in Figure 5 were being run and 15<br />

seconds of the second grinding period had elapsed, the STATUS screen would appear as in<br />

Figure 6.<br />

Figure 6 - Run Screen during grinding<br />

Between any two grinding periods is the COOL DOWN stage, displayed as STATUS: COOL<br />

DOWN.<br />

When the grinding program is complete, the RUN screen will display STATUS: RUN COMPLETE<br />

for approximately 3 seconds, then will revert to STANDBY. TIME REMAINING will display 0:00,<br />

and (in the case of the program displayed in Figure 4) CYCLES REMAINING will display 0 of 3.<br />