If you use a FASTER formalin vaporizer follow this propcedure:<br />
1. after filling the two containers of the vaporizer according to the above-said instructions, the vaporizer can<br />
be connected to the power point inside the cabinet.<br />
2. set the timer at 6 hours.<br />
3. cover the cabinet completely with a polyethylene sheet which is to be carefully sealed at the bottom with<br />
adhesive tape (see enclosed picture 1); also the feeding cable of the cabinet must be sealed as shown in<br />
the enclosed picture 2.<br />
4. start the sterilization cycle by pressing SET and STATUS key together, digit the password (3 times arrow<br />
up + ENTER). All the functions of the cabinet are deactivated except the power point. Now the fans start<br />
working for 30 secs. at intervals of 15 min. for totally 8 hours.<br />
CYCLE: abt. 1 hour of formaldehyde evaporation, 6 hours (time set with the timer) of fumigation process,<br />
1 hour of reagent evaporation. Total time : 8 hours.<br />
5. when the sterilization cycle is over, the message “END STERILIZATION” is displayed<br />
6. to go out of the programme press SET and STATUS key together, digit the password (3 times arrow up +<br />
ENTER)<br />
7. At this point, after wearing protective gloves, protective musk and clothing the polyethylene sheet can be<br />
removed and must be submitted to special waste disposal according to possible local regulations.<br />
8. Aerate profusely the room in order to evacuate completely the residual gas and smell.<br />
This fumigation operation is to be considered as a suggestion.<br />